PR-PD3
Main
External
Dimensions |
Reticle and Mask Particle
Detection System
|
Inspection object |
Particles on reticle/mask with or
without pellicle |
|
Principle |
Laser scattering method |
|
Inspection time |
Approx. 7 min/2 surfaces (5 inch
reticle), Approx. 7 min from inspection start (switch ON) to test
result display for two surface (5 inchs) inspection
* High throughput model specific
to the glass/pellicle (approx. 4 min. for two surface inspection) is
also available. |
|
Detectable particle size |
Pattern surface: 0.5 µm,
Glass surface: 5.0 µm, Pellicle surface: 10.0 µm (* PSL equivalent) |
|
Reticle size |
5, 6 or 7 inch (Please specify
when ordering.) |
|
Reticle cassette (Please specify
when ordering) |
Applicable to the HORIBA standard
cassette for 5, 6, 7 inches as well as stepper cases unique to
individual maker for 5, 6 inches. (Single size of reticle and
cassette handling is the standard. Contact us as to the multiple
cassette handling system.) |
|
Inspection result |
Data mapping display on the FPD |
|
Dimensions |
Approx. 900 (W) x 1350 (D) x 1590
(H) mm
Approx. 35.4 (W) x 53.1 (D) x
62.6 (H) in
(Dimensions will vary depending
on the reticle cassette specifications.) |
|
Installation site |
Clean room or clean booth, class
100 or better |
|
Temperature |
23 ±1°C |
|
Power |
100 to 230 V AC, 7.5 to 5 A,
50/60 Hz, single phase |
|
Vacuum source |
Pressure difference 8.0 x 104 Pa
or more, 50 L/min |
|